Intensive Care Med. 2003, 29(6):955-62Abstract
OBJECTIVES: We evaluated an algorithm for continuous on-line monitoring of alveolar pressure over time in a lung model with lower and upper inflection points and variable resistance ratios and in patients with acute lung injury. The algorithm is based on "static" pressure/volume curves obtained from tracheal pressure measurements under dynamic conditions. DESIGN AND SETTING: Experimental and clinical evaluation of algorithm in a university hospital laboratory and intensive care unit. PATIENTS: Ten patients undergoing postoperative respiratory therapy (feasibility of tracheal measurement) and ten patients with acute lung injury undergoing ventilator treatment (evaluation of algorithm). MEASUREMENTS AND RESULTS: Direct tracheal pressure measurements with a catheter inserted through the endotracheal tube. Comparison of measured alveolar and the dynostatic alveolar pressure vs. time in a lung model with changes in five ventilatory parameters. Examples of clinical monitoring are reported. In the model there was excellent agreement between alveolar pressures obtained by the algorithm, the dynostatic alveolar pressure, and measured alveolar pressure at all ventilator settings. For inspiratory/expiratory resistance ratios between 1:2.1-2.1:1, the dynostatic alveolar pressure was within +/-1.5 cm H(2)O of measured alveolar pressure. In patients the technique for direct tracheal pressure measurement using a catheter inserted through the endotracheal tube functioned satisfactorily with intermittent air flushes for cleansing. CONCLUSIONS: Using a thin tracheal pressure catheter inserted through the endotracheal tube alveolar pressure allows continuous bedside monitoring with ease and precision using the dynostatic algorithm. The method is unaffected by tube and connector geometry or by secretions.Description
To access publisher full text version of this article. Please click on the hyperlink in Additional Links fieldAdditional Links
http://www.springerlink.com/content/95emgpaqjybgvf5yae974a485f413a2113503eed53cd6c53
10.1007/s00134-003-1730-2
Scopus Count
Collections
Related articles
- Evaluation of pressure/volume loops based on intratracheal pressure measurements during dynamic conditions.
- Authors: Kárason S, Søndergaard S, Lundin S, Wiklund J, Stenqvist O
- Issue date: 2000 May
- The dynostatic algorithm accurately calculates alveolar pressure on-line during ventilator treatment in children.
- Authors: Sondergaard S, Kárason S, Hanson A, Nilsson K, Wiklund J, Lundin S, Stenqvist O
- Issue date: 2003 May
- Alveolar inflation during generation of a quasi-static pressure/volume curve in the acutely injured lung.
- Authors: Schiller HJ, Steinberg J, Halter J, McCann U, DaSilva M, Gatto LA, Carney D, Nieman G
- Issue date: 2003 Apr
- Continuous on-line measurements of respiratory system, lung and chest wall mechanics during mechanic ventilation.
- Authors: Kárason S, Søndergaard S, Lundin S, Stenqvist O
- Issue date: 2001 Aug
- Prevention of endotracheal suctioning-induced alveolar derecruitment in acute lung injury.
- Authors: Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L
- Issue date: 2003 May 1