Graefes Arch. Clin. Exp. Ophthalmol. 2011, 249(9):1311-7Abstract
BACKGROUND: The purpose of this study was to test the reliability of vessel diameter measurements with a newly developed retinal oximeter. METHODS: Twelve healthy individuals participated in the study. Retinal images were taken with the Oxymap Retinal Oximeter. Diameters of retinal vessels were measured automatically with the Oxymap Analyzer software. Repeated measurements on the same vessel segments were compared. The automatic measurements were also compared with semi-automatic measurements with a plug-in for the ImageJ software. RESULTS: Variance coefficient (standard deviation/mean) from the repeatability test was 2.8 and 4.0% for first- and second-degree venules and 3.5 and 5.4% for first- and second-degree arterioles, respectively. ImageJ measured larger mean diameters than Oxymap Analyzer in all cases. Means of differences were 5.1 ± 2.2, 2.9 ± 1.3, and 2.7 ± 1.6 pixels for first-, second-, and third-degree venules and 3.1 ± 1.2, 2.7 ± 0.9, and 2.9 ± 1.4 pixels for first-, second-, and third-degree arterioles. CONCLUSIONS: Vessel diameter measurements with the oximeter are repeatable and comparison with an established method demonstrates a relatively stable offset where the standard deviation of the difference is rather small. Different definitions of vessel borders may be the cause of this difference.Description
To access publisher full text version of this article. Please click on the hyperlink in Additional Links field.Additional Links
http://dx.doi.org/10.1007/s00417-011-1680-2Rights
Archived with thanks to Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv für klinische und experimentelle Ophthalmologieae974a485f413a2113503eed53cd6c53
10.1007/s00417-011-1680-2
Scopus Count
Collections
Related articles
- Retinal oximetry images must be standardized: a methodological analysis.
- Authors: Palsson O, Geirsdottir A, Hardarson SH, Olafsdottir OB, Kristjansdottir JV, Stefánsson E
- Issue date: 2012 Apr 2
- Retinal vessel oxygen saturation in healthy individuals.
- Authors: Geirsdottir A, Palsson O, Hardarson SH, Olafsdottir OB, Kristjansdottir JV, Stefánsson E
- Issue date: 2012 Aug 13
- Retinal oximetry and systemic arterial oxygen levels.
- Authors: Eliasdottir TS
- Issue date: 2018 Nov
- Reliability and reproducibility of retinal oxygen saturation measurements using a predefined peri-papillary annulus.
- Authors: Man RE, Kawasaki R, Wu Z, Luu CD, Wang JJ, Wong TY, Lamoureux EL
- Issue date: 2013 Dec
- Retinal oximetry.
- Authors: Hardarson SH
- Issue date: 2013 Mar